欧美国产亚洲另类动漫_在线观看不卡av_91精选在线观看_日韩精品资源二区在线

Products

current position:home page>Products>Analysis Instrument>PST>

Process Moisture Analyzer - Michell Promet EExd

Updated:2025-03-13

Views:3621

  • Brand: PST/Michell
  • Model: Promet EExd
  • Description: Promet EExd Process Moisture Analyzers are hazardous-area rated industrial hygrometer systems for measurement of water vapor content in process gases on natural gas platforms and terminals as well as
  • Feedback
  • WeChat

    QQ

    Online Service

Product Details

Process Moisture Analyzer - Michell Promet EExd Description:

Product Manual Datasheet

Natural Gas Water Dew-Point Analyzer

Features

EExd flameproof certified - IECEx, ATEX, UKCA, cQPSus and TR CU Ex

Complete hazardous area installation

Single- or dual-channel measurement

Moisture content, dew point and pressure

Calibration traceable to NPL (UK) and NIST (USA)

Range -120...+30°Cdp (-184...+86°F), 0.001...30,000 ppmV

Remote global access options

Sour gas compatible

Field calibration maintenance

Promet EExd Process Moisture Analyzers are hazardous-area rated industrial hygrometer systems for measurement of water vapor content in process gases on natural gas platforms and terminals as well as in petrochemical plants and industrial manufacturing facilities.

The Promet EExd is installed directly in the hazardous area, simplifying site installation and retrofits. Promet EExd combines proven Michell moisture sensing technologies with sample conditioning systems designed to provide a reliable and easy-to-operate, multi-channel, on-line instrument for moisture measurement in flammable gases.

Specifications

Channel Configurations Single and two channel
Moisture Parameters Dew point °C and °F
Moisture Content Automatic pressure compensated conversions:
ppmV for natural gas and ideal gas lb/mmscf and mg-m-3 for natural gas Dew point at a pressure input by the user for natural gas and ideal gas.
Calculations for natural gas moisture content based on either IS0 18453 or IGT#8 to customer order preference.
Analysis Pressure barg, MPa and psig
Sensor Technology Michell Ceramic Metal-Oxide Moisture Sensor
Measurement Range -120oC dp...+30oC dp (-184...+86 °F)
1 ppbV...30,000 ppmV
0...250 barg (0...3625 psig)
Calibration Range -100 °C dp...+20 °C dp (-148 °F dp...+68 °F dp)
10 ppbV...23,000 ppmV
Accuracy Dew point: ±1 °C between –59.9 & +20 °C dp (±1.8 °F between -75.8 & 68 °F dp)
Moisture content: ±10 % of reading
Dew point: ±2 °C between –60 & -100 °C dp (±3.6 °F between -76 & -148 °F dp)
Moisture content: ±20 % of reading
Analysis Pressure: ±0.25 % FS.
Measured Resolution 0.1 °C (0.18 °F): –80...+20 °C dp (-112...+68 °F)
1 °C (1.8 °F): –100...-80 °C dp (-148...-112 °F)
Displayed Resolution Dew point: 0.1 °C (0.18 °F)
Moisture content: autoscale, 5 digits
MPa and BarG: 0.1, psig: 1.0
Resolution 0.1 °C between +20 °C dp and -80 °C dp (0.18 °F between +68 and -112 °F dp)
1 °C between -80 °C dp and -100 °C dp (1.8 °F between -112 and -148 °F dp)
Temperature Coefficient Algorithm compensation (-20...+50 °C)
HMI: Keyboard Interface Capacitive touch-screen through glass
Display Vacuum fluorescent
Datalogging A rolling maximum of 150 data logs is available.
Each log records time, date, moisture and pressure values for each channel.
5 minutes minimum and 60 minutes maximum logging intervals can be set by the user
Communications and Output Two non-isolated 4...20 mA per measurement channel. 500 Ω maximum load. Range and parameter settable by user.
Modbus RTU @ 9600 baud-rate
Alarms: two volt free contacts per channel; one process value and one instrument status. Also available via Modbus communications
Main Unit Enclosure Internal temperature control for condensation protection and stable analysis conditions
Type Explosionproof NEC/Flameproof EExd
Construction Cast copper-free aluminium
Finish Chromate primer, Polyester P9010 powder coated (black). Meets BS3900
Operating Pressure 30...138 barg (435...2000 psig)*
Sample Connections 1/8” NPT (F)
Sample Flow Rate 1 Nl/min (Nm3/hr) (2.1 scfh)
Sample Flow Alarm Optional
Power Supply 90...260 V AC, 50/60 Hz, 180 W Main Unit
Operating Environment Indoor/Outdoor -20 °C...+60 °C (-4...+140 °F)
Max 95 % rh. Shaded location
Certification
Hazardous Area Certification ATEX/UKCA: II 2 G Ex db IIB+H2 Gb
T5 (Tamb -40 °C...+44 °C) – 60 barg max.
T4 (Tamb -40 °C...+60 °C) – 60 barg max.
T3 (Tamb -40 °C...+60 °C) – 138 barg max.

IECEx: Ex db IIB+H2 Gb
T5 (Tamb -40 °C...+44 °C) – 60 barg max.
T4 (Tamb -40 °C...+60 °C) – 60 barg max.
T3 (Tamb -40 °C...+60 °C) – 138 barg max.

cQPSus: Class I, Division 1
Groups B, C & D
T5 (Tamb -25 °C...+44 °C) – 60 barg max.
T4 (Tamb -25 °C...+60 °C) – 60 barg max.
T3 (Tamb -25 °C...+60 °C) – 138 barg max.

CL I ZONE 1: Ex db IIB+H2 Gb
AEx db IIB+H2 Gb
T5 (Tamb -40 °C...+44 °C) – 60 barg max.
T4 (Tamb -40 °C...+60 °C) _ 60 barg max.
T3 (Tamb -40 °C...+60 °C) – 138 barg max.
Additional TRCU 012 (EAC), PESO (India), KCS (Korea), Ukraine Ex
Pattern Approval GOST-R, GOST-K
*Other variants available to customer specific order – consult Michell Instruments

Suitable Applications

Natural gas glycol dehydration

Offshore export pipeline natural gas

Transmission pipeline monitoring

Hydrogen production, storage and transportation including natural gas injection

LNG production

Ethylene feed stock to polymer plant

Speciality gas production

WeChat

Customer Service QQ

Customer Hotline:

010-52867770
0l0-52867771
l3811111452
l7896OO5796

Technical Supports

0l0-52867774
l7896OO5796
info@m.jidi9.com
主站蜘蛛池模板: 日本不卡高清视频一区| 久久精品最新地址| 久久亚洲国产成人| 国产日韩第一页v| 欧美欧美一区二区| 色综合五月天导航| 国产精品毛片va一区二区三区 | 日韩a在线播放| 97精品伊人久久久大香线蕉| 精品中文字幕在线2019| 久久99中文字幕| 欧美 日韩 国产精品| 日韩中文字幕在线视频| 亚洲一区在线直播| 在线视频一区观看| 亚洲欧洲免费无码| 中文字幕日韩一区二区三区| 91国内揄拍国内精品对白| 99在线视频免费观看| 国产成人av一区二区三区| 国产精品国产亚洲精品看不卡 | 精品国产综合| 国产自产在线视频一区| 国产在线精品91| 国产精品一区二区三区观看| 国产综合在线观看视频| 精品国产一区二区三区久久狼黑人 | 欧美激情国产日韩精品一区18| 日韩av资源在线| 日本韩国在线不卡| 欧美韩国日本精品一区二区三区| 久久免费少妇高潮久久精品99| 久久精品视频中文字幕| 国产精品一级久久久| 国产精品美女www爽爽爽视频| 91国产在线精品 | 亚洲精品在线免费看| 日韩三级在线播放| 萌白酱国产一区二区| 黄色片视频在线播放| 国产精品自产拍在线观看中文|